Research Article: | Validation of a headspace gas chromatography-flame ionization detector (HS GC-FID) method for determination of terpinen-4-ol in volatile oil from Zingiber montanum rhizome |
Author: | Nirawan Kitprapiumpon, Shinnawat Saengungsumalee, Thitaree Theerachayanan, and Surapote Waongyai |
Email: | shinnawat@siam.edu |
Department/Faculty: | Faculty of Pharmacy, Siam University, Bangkok 10160 |
Published/แหล่งเผยแพร่: | RSU International Research Conference 2021 on Science and Technology |
Citation
Nirawan Kitprapiumpon, Shinnawat Saengungsumalee, Thitaree Theerachayanan, and Surapote Waongyai. (2021). Validation of a headspace gas chromatography-flame ionization detector (HS GC-FID) method for determination of terpinen-4-ol in volatile oil from Zingiber montanum rhizome. In The 6th International Conference on Sciences and Technology (RSUSCI-2021) (pp.27-36). Pathum Thani: Rangsit University.
ABSTRACT
This study aimed to develop a headspace GC-FID method for the analysis of terpinen-4-ol in Zingiber montanum (Plai) essential oil. The volatile oil was extracted from fresh Z. montanum rhizomes by steam distillation. The amount of terpinen-4-ol was determined by a headspace gas chromatography-flame ionization detector (HS GC-FID) method using benzyl alcohol as an internal standard. The developed method was validated according to the parameters recommended by the International Conference on Harmonization (ICH). Linearity was accessed across the concentration range of 2.5 – 15 µL/mL. The plot of the peak area ratio versus the concentration provided a good linear of this method with a correlation coefficient (r2) of 0.9989. The LOD and LOQ were found to be 0.20 and 2.5 µg/mL, respectively, which indicated a high sensitivity of the method. The instrument (n = 6), intra-day (n = 9), and inter-day (n = 3) precision as indicated by %RSD were 1.77, 1.73, and 1.77 %, respectively. The recoveries at 3 different levels of terpinen-4-ol were between 100.23 to 100.85%. The content of terpinen-4-ol in Plai essential oil was 15.53 ± 0.27 % v/v. The developed method can be used in routine analysis of Plai essential oil and its products in both cosmetic and pharmaceutical industries.
Keywords: Terpinen-4-ol, Zingiber montanum, Plai, Gas chromatography, Validation
Validation of a headspace gas chromatography-flame ionization detector (HS GC-FID) method for determination of terpinen-4-ol in volatile oil from Zingiber montanum rhizome
Faculty of Pharmacy, Siam University, Bangkok, Thailand
Related
- Thromboprophylaxis in Multiple Myeloma Patients Receiving Immunomodulatory Drugs
- The Perspective of Stakeholders towards the Problems and Solutions of Health Products Available Through Social Commerce in Thailand
- Prognostic Value of Serum Procalcitonin level for the Diagnosis of Bacterial Infections in Critically-ill Patients
- Formulation and in-vitro evaluation of fast dissolving tablets using superdisintegrant blend with effervescent materia
- Doripenem dosing regimens in Asian critically ill patients with continuous real replacement therapy
- Doripenem dosing in Asian critically ill patients with continuous renal replacement therapy
- Design and characterisation of electrospun shellac-polyvinylpyrrolidone blended micro/nanofibres loaded with monolaurin for application in wound healing